Press âApplyâ button to start new probe optimization (Figure 3.7).
</p>
</ul>
[[Image:neoscanfig_3_15.png|thumb|right|400px300px|<i><b>Figure 3.15</b>: Dialog window for initiating the optimization process.</i>]] A user is presented with a dialog window shown in Figure 3.15, giving an option to initiates the optimization process â which will take about 20 minutes to be completed â or skip the optimization by choosing âNo.â Click on âYesâ for optimization. The system will remind a user to prepare an optimization structure. Align crystal field sensitivity direction to optimization structure field properly and press âContinueâ.
When the optimization process starts, [[NeoScan]] Optimization Utility window will pop up, and the detailed optimization parameters are displayed to a user (Figure 3.16): Table 3.1 lists the main parameters used in optimization process with their default values.