NeoScan Manual Part B: Measurement Preparations

1 byte removed, 23:15, 6 March 2017
/* System Monitor Page */
To start a new probe optimization press “Apply” button.
</p>
</li></ol>
If the polarization is maintained, the total return power and the polarization power will lie within the specified optimized range (the default value is ±20%). In this case, “Optimization Indicator LED” becomes green. Otherwise, it start blinking and the information panel indicates: “Polarization is not maintained: Optimization is required.” A user can change the threshold from the box labeled “Auto Trigger Margin” (see Figures 3.7 and 3.8). If the power difference is greater than 20%, the polarization state is found not to be maintained, therefore, the user will be prompted with a dialog window giving an opportunity to initiates the optimization process or skip the optimization by choosing “No.”
</ul>
 
</li>
</ol>
==== Optimization Settings Page ====